Skip to main content

Table 4 Continuous episodes of temperature excursion above 8 °C in cold chain devices

From: Temperature integrity and exposure of vaccines to suboptimal temperatures in cold chain devices at different levels in three states of India

  

Total observation perioda (hours)

Exposure to temperature > 8 °C

Total durationa (hours)

Total continuous episodes durationa (hours)

Duration of continuous episodesb

≤1 h

> 1–9 h

≥10 h

State and Divisions

Bihar

4107.5

302.7

14.5

4.25

8

2.25

Gujarat

1395

26

7

7

0

0

Kerala

860.5

3.5

0.2

0

0.25

0

Total

6363

332.2

21.7

11.25

8.25

2.25

District

Bihar

5163.7

1318.7

2.2

0

0.5

1.75

Gujarat

3228.7

14.5

0.5

0.25

0

0.25

Kerala

1914.5

0.5

0.2

0.25

0

0

Total

10,307

1333.7

3

0.5

0.5

2

Sub-district

Bihar

23,728.2

1846.2

28.5

8.75

11.75

8

Gujarat

12,115.2

186

6.5

4

1.2

1.2

Kerala

6526.2

182.7

2.2

0.5

0

1.7

Total

42,369.7

2215

38.2

13.5

12.7

12

WICs

Bihar

3483.2

299.2

13.7

3.7

8

2

Gujarat

1036.7

26

7

7

0

0

Kerala

718.5

3.5

0.2

0

0.2

0

Total

5238.5

328.7

21

10.7

8.2

2

ILRs

Bihar

28,452.2

2578

27.2

9

11.2

7.5

Gujarat

15,702.2

200.5

7.5

4.5

1.2

1.5

Kerala

8582.7

183.2

2.5

0.7

0

1.7

Total

52,737.2

2961.7

37.2

14.2

12.5

10.7

Other devicesc

Bihar

1064

590.5

4.75

0.7

0.7

3.2

Total

1064

590.5

4.7

0.7

0.7

3.2

Pooled all levels

Bihar

32,999.5

3467.7

46.2

13.5

20

12.7

Gujarat

16,739

226.5

14.5

11.5

1.2

1.5

Kerala

9301.2

186.7

2.7

0.7

0.2

1.7

Total

59,039.7

3881

63.2

25.7

21.5

16

  1. WIC Walk-in cooler, ILR Ice lined refrigerator
  2. a Duration in hours
  3. b A continuous episode above 8 °C was counted when readings were continuously above 8 °C for at least 45 min or more
  4. c Other devices (domestic refrigerator and cold boxes) were used for vaccine storage in Bihar only at the time of the study